Exponent Connect Firmware version history
9th November 2026
- Added task-stall monitoring for communications with the Motor processor, the Control Panel and USB mass storage. When enabled, the watchdog resets the system if a monitored task stops reporting progress for two minutes.
- Added task-stall details to the startup log, System Status reset diagnostics and a Control Panel alert after reboot.
- Improved standalone test startup checks so a test is not started if the required results or filename buffers cannot be allocated.
- Fixed malformed or oversized Control Panel event handling and improved resource-request buffer cleanup when memory is low or the upload queue is full.
- Increased the Motor communications task stack and moved several task stacks and file path buffers to external RAM to reduce internal memory pressure and prevent stack overflow during file operations.
- Updated operating system checks to support task stacks and synchronisation objects in external RAM.
- Improved Control Panel graph data transfer retries and prevented simultaneous display requests from interfering with acknowledgements or replies.
- The LCA calibration completion message now displays the calibrated gain on both Control Panel layouts.
- Fixed startup update prompts that could wait indefinitely when no display, PC or web connection was available to answer them.
- Improved handling of corrupt or truncated update packages, FPGA images, user profiles, probe databases and language files, including cleanup of incomplete files and memory allocation failure handling.
- Exponent file downloads now use temporary files and replace the destination only after successful completion, with backup and restore handling if replacement fails.
- Updated HTTPS to use Mbed TLS, with ECC P-256 certificate generation on a separate task and improved TLS memory allocation and diagnostics.
- Added hardware AES acceleration for HTTPS using the processor’s cryptographic accelerator.
- Limited the initial DHCP request to 20 seconds and added fallback to the configured static IP address if DHCP fails.
- Fixed invalid network settings that enabled both the DHCP client and DHCP server at the same time.
- Improved mDNS hostname resolution for Windows and browser clients, including mixed-case names and address-query handling.
- Added PC USB/UART communications performance statistics and an on-device AES-GCM benchmark to the diagnostic shell.
- Added regression checks for task-stall recovery, Control Panel communications, mDNS, Ethernet driver handling, TLS and communications performance monitoring.
30th September 2026
- Added laser calibration functions and improved quick-position control, TSI calibration screens, and TSI buffer handling.
- Added system status and memory monitoring, and improved Modbus TCP and reset diagnostics.
- Improved mDNS and session handling; disabled FTP.
- Improved File management and web uploads, including progress/error handling and copying hidden files; fixed large NIO buffer writes and stopped sending test data after a connection is lost.
- Improved memory allocation, task stack sizing, and USB handling.
- Fixed some issues with importing and exporting results.
30th June 2026
- Improved language flag handling and missing-resource upload tracking for the Big TSI display.
- Added robustness improvements in error handling and memory allocation checks across comms/event paths.
- Enhanced test run setup and standalone test flow handling.
- Updated Live Readings behaviour, including button handling and Volscan-specific display updates.
- Refactored Big Display tab handling and improved sequence variable transfer behaviour.
- Added a new Run tab on Big Display: "Last Test".
- Added per-project configuration for Last Test screen values (up to 4 entries), including "None" selection support.
- Added backward-compatible project save/load updates to persist Last Test display settings.
- Added dedicated user option to auto-open Last Test at end of standalone test, while preserving existing Results-at-end behaviour as fallback.
- Added result limit colouring to Last Test big-value fields using project min/max checks (consistent with Results table behaviour).
29th April 2026
- Fix Break Detect Level settings.
22nd April 2026
- Added a method to reset chassis information via a chassis.bin file.
9th March 2026
- Check for production test Guard so it is not permanently remembered.
4th March 2026
- Added Next button and delay to startup log.
- Request free storage space to be calculated when firmware information retrieved.
24th February 2026
- Clearing the Error Flag now clears the End Test Flags as well so errors are not reported at the end of the test.
24th February 2026
- Fix an issue with using IF statements in the Abort sequence.
9th February 2026
- Fix an issue where Height Calibration would fail after the CP Stop button had been pressed.
29th January 2026
- Added extra engineer functions.
24th December 2025
- CP reset button now returns if the sequence clears the RESET flag so it can be used again during the same test.
- Extended the length of the status message display space when a test is running.
17th December 2025
- Added system warning check for user gain being more than 10% different to the factory gain.
9th December 2025 – Official release of version 4.xxxxx firmware.
- Support for the new 100kg Texture Analyser.
- Updated TSI interface.
- Added support for the new Control Panel (TSI replacement).
- Added Touch To Move feature when using the Control Panel.
- Fixed an issue with the RS422 communications.
- Enhanced USB keyboard usage when using the Control Panel.
- Added more languages:
- TSI: Czech, English, Finnish, French, German, Hungarian, Indonesian, Italian, Lithuanian, Polish, Portuguese (Brazil), Portuguese, Romanian, Spanish, Turkish.
- Control Panel: Chinese, Czech, English, Finnish, French, German, Greek, Hungarian, Indonesian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese, Romanian, Russian, Serbian, Spanish, Thai, Turkish, Vietnamese.
6th September 2024
- Fixes issue with status message slow down after a dialogue has been displayed.
2nd September 2024
- Fixes for T.A. sequence format string command.
2nd September 2024
- Fixes for reboot issues in last update.
11th July 2024
- Fixes for Rate Break detect.
29th May 2024
- Fixes for TSI not redrawing after it has updated the firmware.
23rd May 2024
- Fixes for VS screen layout.
11th April 2024
- Extended timeout when wait for an ACK from the TSI to prevent duplicate messsages.
8th April 2024
- Fix showing/hiding Run button.
11th March 2024
- Fix issue with new Home screen recursion and Live Position update.
11th March 2024
- Fix issue with non-ASCII characters in company name causing issue with EEPROM messages.
5th March 2024
- Set new Live Readings as the default startup page.
5th March 2024
- Fix TSI issue in new Live readings page
5th March 2024
- Fix TSI issue in new Live readings page
5th March 2024
- Add option to send data to I2C peripheral via the 485 commands by setting a negative ID.
27th February 2024
- Increase timeout on RS422 reception to fix errors communicating with the RAED.
Older versions
3.00472 10th January 2023
- Fix issue with moving the probe via the USB connection.
3.00473 2nd February 2023
- Added a diagnostic command for production.
3.00474 5th April 2023
- Fix an error message issue.
3.00475 14th April 2023
- Added new motor controller type to list.
3.00476 1st June 2023
- Improve accuracy of laser reading rotational sync.
3.00477 3rd August 2023
- Added more abort routes to long processes.
3.00478 22nd August 2023
- Added more abort routes to long processes.
3.00479 26th October 2023
- More space added for Run and Cycle numbers in the results list.
3.00480 28th November 2023
- Added TSI Display Theme support.
3.00481 12th December 2023
- Fix export of tab results.
3.00482 12th December 2023
- Fix issue with configuration data after firmware update
3.00430 11th January 2022
- Enlarged area for big live readings values.
- Changed TSI probe move method as tilt sensors are now unobtainable.
3.00431 1st February 2022
- Increase motor acceleration timeout for slow RPS.
3.00432 14th February 2022
- Fix rotary deceleration issue.
3.00433 17th February 2022
- Enhanced VS calibrate height routine to work with new verification rod.
- Changed Video / AED trigger signal to keep it active while data is captured.
3.00434 17th February 2022
- Allow 32 bit MODBUS register addresses in the device log map.
3.00435 22nd February 2022
- Increase allowance for Mid-Point value.
- Restrict reporting of blocked / failed network connections to once per 10 seconds per IP.
3.00436 22nd February 2022
- Fixed regression to old video capture box not recognising the new trigger message.
3.00437 3rd March 2022
- Don't spin the product for product check if the product is already spinning.
3.00438 8th March 2022
- Reduce load cell over and under limits and widen mid-point to allow for tolerance of new load cells
3.00439 10th March 2022
- Reset VS rotatory position at the start of each scan.
- Increased top pin diameter for alignment wizard.
3.00440 17th March 2022
- Show the new result on the TSI when the results list is on view and the test finishes.
- Fix possibility of double FAST + RUN button detection when starting a test.
3.00441 31st March 2022
- Added option to move to Force / Position by taping on the force / position field in the Big Live Readings screen.
3.00442 5th April 2022
- Set Spin Speed to 1.0 RPS during alignment wizard.
3.00443 20th April 2022
- Extended top of product search timeout
3.00448 22nd April 2022
- Fix for new AED trigger signal
3.00449 26th April 2022
- Ensure new AED trigger signal is switch off when a test is aborted.
3.00450 19th May 2022
- Added extra information message to the height calibration routine
3.00451 19th May 2022
- Message displayed on the TSI when a keypad button is pressed if the keypad is locked.
3.00455 27th May 2022
- Added option to update TSI Probe Height Calibration values after the height is calibrated.
3.00456 7th June 2022
- Validate Frame Deflection Calibration user name.
- Reduce limits for VSP Calibration Rod detection.
3.00457 16th June 2022
- Validate stored user names.
3.00459 23rd June 2022
- Fixes to Display Message when working with Exponent Connect.
3.00460 30th June 2022
- Fixes to Display Message when working with Exponent Connect and no TSI fitted.
3.00462 20th July 2022
- Fix distance offset firmware information.
3.00464 17th August 2022
- Now returns a more detailed error code when settings parameters
3.00465 27th September 2022
- Added more description for calibrate laser settings.
3.00466 13th October 2022
- Added option to wait for Move Probe comm's command to finish.
- Split "Top or Bottom" stop return error message to show which stop was active.
- Fixed race condition with stop processing.
3.00467 3rd November 2022
- Update VS alignment wizard.
3.00468 10th November 2022
- Update VS alignment wizard.
3.00469 15th December 2022
- Increase timeouts to cope with very low spin speeds.
3.00470 20th December 2022
- Fix issue with starting a test via Ethernet connection.
- Fix issue with moving the probe via the Ethernet connection.
3.00471 22nd December 2022
- Fix issue with moving the probe via the USB connection.
3.00397 14th January 2021
- Fixed issue where no results were displayed for new projects.
3.00399 15th February 2021
- Add support for internal 422 port
3.00400 22nd February 2021
- Ignore remote when running test via Batch mode
3.00401 2nd March 2021
- Added User name that performed the laser offset calibration.
3.00402 2nd March 2021
- Added "Special" load cell model option.
3.00403 3rd March 2021
- Added diagnostic option to scan when the bottom stop is active.
3.00405 11th March 2021
- Added new a break detection mode that triggers when the force drops below X% of the maximum measured force. Requires Exponent Connect 8,0,4,1 or later
3.00406 21st April 2021
- Added Volscan laser serial number to the list of information data.
3.00407
- Added Location to firmware information.
- Changed the Volscan laser stop alignment message for new procedure.
- Added motor dwell setting to auto options.
3.00408 20th May 2021
- Added memory capacity parameter.
3.00409 24th May 2021
- Fixes for DR DIS pressure checks.
- Added Pressure captions to replace Force for DR DIS.
3.00410 4th June 2021
- Added an option to the COMS API to wait for a new rotational average laser reading.
3.00411 29th June 2021
- Added an option to update peripheral firmware.
3.00412 29th July 2021
- Fixed gain check for new DrDis units
3.00413 18th August 2021
- Added Large Live readings requires TSI v94.
- Added sign on / off LAN broadcast message.
3.00414 20th August 2021
- Fixed Get485Data32 command.
- Increased buffer for sign on / off LAN broadcast message to accommodate extra information.
3.00415 26th August 2021
- Fixed alignment wizard door override.
- Added option to verify rotary axis start up.
3.00416 3rd September 2021
- Fixed issue with value update on alignment wizard door override.
- Fixed marginal timing on rotary startup check.
- Added option to move up / done on top pin wizard to cope with both old and new platforms.
3.00417 6th September 2021
- Fixes to Volscan factory setup options.
3.00418 7th September 2021
- Fixes to Volscan factory setup options.
- Fix Volscan Rotary units text on TSI.
3.00419 9th September 2021
- Fix DrDis capacity setting.
3.00420 15th September 2021
- Added diagnostic command to set the laser offset
3.00421 5th October 2021
- Fix prompt to repair chassis / mca data.
3.00422 14th October 2021
- Added Volscan bottom pin alignment wizard.
3.00423 28th October 2021
- Added Volscan height offset calibration temperature
3.00424 29th October 2021
- Fix Volscan height offset calibration temperature.
3.00425 10th November 2021
- Product check at top stop now user minimum radius value.
3.00426 30th November 2021
- Enhanced VS slow Spin up to allow for faster rotation speeds.
3.00427 9th December 2021
- Made 5 RPS the default maximum spin speed.
3.00428 16th December 2021
- Fixed Constant Strain Rate sequence command.
- Fixed Bubble Strain Rate sequence command.
3.00429 23rd December 2021
- Fix issue with starting the probe move when bend compensation is switched on.
3.00357 9th January 2020
- Added Laser On / Off command to PC interface.
- Fix bad load cell serial number when zero.
3.00358 13th January 2020
- Added Disable Rotary Motor Option to Volscan.
3.00368 3rd February 2020
- Changed TSI Close button to operate on button up so it doesn't effect the next screen.
- Added "Force Calibration Complete" message to the Message log.
- Added code to repair MCA settings from a 'mca.bin' file supplied by SMS.
- Moved error detection / reporting functions to a different thread so the transmission of status message is not disturbed.
- Added facility to test Video Capture box trigger input by pressing STOP and RESET to simulate a test trigger point.
3.00369 10th February 2020
- Fixed MODBUS-TCP/IP crash when too many registers were requested.
3.00373 20th February 2020
- Added Load Cell Model to the firmware information.
- Fixed Restore option not saving recalled user list.
- Fixed a timing issue when starting the motor immediately after hitting a stop.
- Added option to wait for filter to settle when switching channels.
3.00374 25th February 2020
- Enhance laser error code for the Volscan.
3.00378 5th March 2020
- Check laser is returning good values after turning it on.
3.00381 2nd July 2020
- Fixed timeout when spinning up a heavy mass.
- Fixed issue with laser switching off and on when Scanner Utilities window was using Continuous rotation.
3.00385 8th August 2020
- Add option to use , for decimal point.
- .CTF Test file now saves PPS, batch ID and probe name. Requires Exponent v8,0,0,20 to read the new file format.
3.00392 14th September 2020
- Add option for 2 point calibration on Volscan.
- Check for valid gain after calibration.
- Added beep to T.A. trigger data capture on. Option is switched on in the Admin - System Settings on the TSI.
3.00393 18th September 2020
- The TSI results are now refreshed at the end of the test when break detection is enabled.
3.00394 21st September 2020
- Fixed check for product function on Volscan.
3.00396 2nd November 2020
- Add option for L6N load cell.
3.00277 17 May 2019
- Updates for production setup
- Fixed scroll position issue when returning to various screens after opening a sub-screen.
- Added TA Sequence and TA Settings modified date and time to TA Settings screen.
- Added section titles to Firmware Information screen.
- Improved compatibility with Firefox browser.
- Split Chassis and Systems Information into separate sections on Firmware Information screen.
- Added feature to show both chassis and backup values when they are different to Firmware Information screen.
- Extra System Warnings are detected.
- Added prompts before fixing chassis/cache EEPROMs after swapping boards
- Added options to the project for maximum / minimum result limits. Results below the minimum are displayed in blue and results above the maximum are displayed in red.
- Fixed file system issue when reading data from some USB drives that have no standard partition.
- Safety warning displayed on TSI when instrument switched on.
- Front panel COMMS LED now comes on when a connection is established over USB or Ethernet and flashes upon data transfer. (Requires FPGA revision 28/11/2018 22:20:59 or later)
- Final value now marked at target position in line with old instruments. (Requires FPGA revision 28/11/2018 22:20:59 or later)
- Fixed possible motor stall when arm moved, stopped and moved again in very quick succession.
- Fixed issue with move probe command from Exponent hangs when used over Ethernet and no safety cabinet fitted.
- The "Waiting for run button" message can now be aborted from Exponent Connect (>= V7.0.4.0) and is cleared when the Run button is pressed.
- The results can now be configured by an Admin to show values relative to the start of the cycle.
- The exported TAB result file now includes the values at the start of each cycle.
- The small graph can now be configured to show absolute or relative data and the distance can be inverted.
- Motor enable delayed at start up to allow PSU to stabilise before drawing power for the motor. Required for HD models to improve switch on reliability.
- Fixed issue with moving to a lower force than already applied to the probe.
- Added Target and Speed to the text in the Quick Position selection drop list.
- Web results list now recovers scroll position when the view is updated.
- Fixed issue where some TSI screen were not closed that needed to be closed when Exponent connected.
- The force on the simple Live Readings screen is now averaged for a more stable reading. Long press on the screen to change to the complete list and the force is direct as before.
- Some screens now use frames to group related items for a cleaner look. (requires TSI firmware 85 or later)
- Added options to the General Security to enable / disable 'User level' users access to various features.
- Added options for Backup and Restore of system settings on a USB stick.
- Added options to display graphic buttons in place of some text buttons. (requires TSI firmware 85 or later)
- Added Polski to language options.
- Added option to move the probe by tilting the screen. (requires TSI firmware 85 or later)
- Added help button to show T.A. keypad shortcuts.
- Added QR code to allow easier connection of mobile devices that are on the same network. (requires TSI firmware 85 or later)
- Added QR code to allow easier navigation to the registration page, with some parameters pre-filled, via a mobile device. (requires TSI firmware 85 or later)
- Added System Warnings screen to view any errors that were detected during boot.
- Many other stability fixes have been included.
3.00280 30th May 2019
- Changed drive strength on A/D data lines to fix intermittent start up issue on some boards.
3.00281 3rd June 2019
- Results screen now scrolls to the bottom when opened so the latest result is in view.
- Fixed issue where Results screen would jump back to the top after a message box was displayed.
- The check boxes in the File Manager screen are now cleared when certain options are changed.
3.00282 12th June 2019
- Change debounce timers back to be compatible with old FPGA.
- Added "Height Calibration Complete" message to bubble and message log.
3.00300 8th August 2019
- Fixed break detect rate mode.
- Debounce timer values now based on FPGA number.
- Extra verification on messages passed between the two processors.
3.00301 13th August 2019
- Fetching Parameters using the new method ignores the mezzanine type.
3.00304 15th August 2019
- Increased timeout for bend calibration and added status messages.
3.00311 29th August 2019
- Fixed setting chassis board serial number on Volscan
- Fixed Auto values for Volscan motor setup
- Updated Volscan Alignment Wizard.
- Changed I2C speed to be slower on Volscan due to extra cable length.
- Fixed "Set Target Aux" sequence command.
3.00312 4th September 2019
- Fixed some timeouts that were too short
- Comm's messages are now rejected when the queue is full instead of wait for a slot that can cause the software to timeout.
- Fixed a race condition if two move commands were sent in very quick succession. The second request is now rejected instead of timing out.
- Fixed a race condition between the two microprocessors at start up that could cause a delay.
3.00317 3rd October 2019
- Fixed issue transferring commands to motor processor.
- Changed Volscan laser configuration for clearer image.
3.00348 6th December 2019
- Re-write of postbox between mains processors to improve speed.
- Enhanced EEPROM test function of Motor console port.
- Add CPU load indication for both processors to TSI and Web Live readings.
- Fixed issue where laser calibration was lost when the instrument was switched off.
- Added a 0.5s delay between energising and starting the motor.
- Fixed calibration issue with DRDis.
- Extended timeout for probe height calibration.
- Changed default motor parameters for Volscan.
3.00355 18th December 2019
- Fixed some Volscan production diagnostics items.
- Fixed FPGA postbox race condition.
- Fixed DR DIS calibration issue.
3.00162 25th April 2018
- First release for production
3.00163 26th April 2018
- Fixed issue with project selection in TSI File manager
- Improved failure detection on uploading a file to the TSI and offer an option to try again.
3.00164 26th April 2018
- Fixed issue with updating TSI firmware introduced in 3.00163.
3.00165 27th April 2018
- Fixed issue with putting a load cell back into an old XTPlus after being in a Connect.
- Add Opto Input status to the Live Readings on the TSI.
- Added Opto Output toggles the Diagnostics screen.
3.00166 30th April 2018
- Added ID numbers that were missing from parameters in the Firmware Information.
- Added parameters that were in missing in grouped lists.
- Added Opto Input status to the Diagnostics screen.
- Added LastFile option to the Exponent Connect download command to bypass the delay to wait for more files.
3.00167 1st May 2018
- Added facility to send, delete and check if files exist on the TA.
- Protected the TA Projects folder to prevent accidental deletion.
3.00168 11th May 2018
- Added Total Number of Test Runs and Total Time Powered On.
- Added more support for future VolScan use.
3.00169–3.00178 31st May 2018
- Added more support for future VolScan use.
- Added audible feedback for TSI touch response and message boxes.
- Added option for user position offset so probe can be calibrated of a known surface that is not zero height.
- Added a queue system for inter-processor tasks for better performance.
- Extended number of characters for registration code on TSI to 50.
- Extended number of characters for company name on TSI to 50.
- Extended number of characters for company name on HTML interface to 99.
- Added extra characters to keypad for company name screen on TSI.
- Added support for & symbol in HTML interface fields.
3.00180 11th June 2018
- The Show Warnings screen now shows if the Real Time Clock is not working.
- Most floating point information is now shown to 6 digit precision.
- Fixed missing resource error for Sample Source.
- Fixed issue with the "(Var X) = {16bit} data from 422 register (Var Y)" sequence command.
3.00181 13th June 2018
- The Network screen now shows the network cable connection status and refreshes if the network connection changes.
- Fixed Web interface issue if a test is run while at the login screen.
- Fixed handling of invalid sequence commands so test now aborts to prevent unintentional actions.
- Invalid sequence commands are now reported with the line number and command name.
- Added extra TSI status message for network reject information.
- Removed a lot of information from Live readings screen so the essential information is easier to see. A long press on a blank area of the TSI of the Live Readings screen will bring them back.
3.00182 20th June 2018
- Improved start-up speed.
- Fixed intermittent corruption on the TSI after performing a touch calibration.
- Fixed issue where corrupted .pac file could slow the system down
- Added more boot up progress messages.
3.00183 21st June 2018
- Fixed Tare button position on Live Readings Screen.
- Added Pitch and Roll to advanced live readings which is calculated using the on-board accelerometer.
3.00184 5th July 2018
- Fixed Height Calibration not reporting errors correctly.
- Stop button now aborts Height Calibration
- Added check for user calibration gain against factory gain
3.00185 9th July 2018
- Added Multi-Point calibration date / time, user and Loadcell serial number to firmware information.
3.00187 22 August 2018
- Fixed issue with Express instruments reporting a bad Expansion EEPROM.