Force Range
+/- 5kg +/-30kg +/-50kg
Force Resolution
0.1gm 0.1gm 0.1gm
Loadcells
Directly interchangeable by the user. All loadcells store their unique calibration and identification information in 'onboard' non-volatile memory
Speed Range
0.01 - 40mm-sec
Speed Accuracy
Better than 0.1%
Range Setting
0.001-295mm
Range Resolution
0.001mm
Data Channels
Filtered force at 20 bit
Distance at 24 bit
Unfiltered force at 16 bit
Two linear analogue inputs (range +/- 4.5v at 16 bit) or PT100 temperature probe inputs (range -50°C to +250°C)
Biphase digital encoder input at 24 bit suitable for any compatible linear or rotary extensometer.Data Acquisition Rate
Up to 500 points per second (pps) for each data channel.
Filtered Force
Oversampled at 8000 samples per second and digitally filtered to 500 pps at 20 bit resolution.
External Instrumentation Channels
Four channels of RS485 using an industry standard MODBUS protocol. Each channel logs at one sample per ten seconds at 16 bit and is suitable for external sensing of temperature, humidity etc.
Operating Temperature
0 - 40°C
Operating Environment
Laboratory conditions. Dust and splash resistant.
Net weight
16.2kg
PC Interface
Interface to PC through a standard RS232 serial port at 115200 BAUD.
Power supply
Universal mains input voltage.
Firmware updates
FLASH update of firmware via PC.
Copyright © 1996-2008 Stable Micro Systems Ltd