The system is supplied with a factory force calibration (which can be restored at any time). However, if you wish to calibrate at a specific weight, you can perform the following procedure:

 
FORCE

Calibration window

1: From the tool bar menu, click on T.A. > Calibrate > Calibrate Force.

Loadcell window

2: Select either Factory or User, as required. The installed loadcell capacity is identified.

Calibration weight window

3: Select the calibration weight of your choice.

Force calibration successful

4: This window should be displayed. Remove the weight from the calibration platform.

PROBE HEIGHT

Only required if you wish to: 
a: measure in % strain;
b: record the height of the product during the test;
c: use a button trigger;
d: start each test from the same position;
e: pre-set probe positions in the Move Probe window.

Calibration window

1: From the tool bar menu, click on T.A. > Calibrate > Calibrate Height.

Probe height window

2: Select the Return Distance and the Speed you wish the probe to move to this once it has touched the contact surface (0mm).

Probe height calibration successful

3: If probe height calibration has been successful, this window will be displayed.